Technical Specification



BlendAlyser NIR 0018 mems
Spectral range1350 – 1800nm
Optical resolution (FWHM)1 – 5 nm
Wavelength accuracy 0.025nm
Wavelength reproducibility 0.01nm
Nonlinearity<5%
Noise - High-Fluxbetter than 0.3 x 10-3 AU
Noise – Low Fluxbetter than 1 x 10-3 AU
Sampling interfaceNon contact through a window
Measurement spot size25mm approx.
Working distanceDepending on thickness of the Sapphire window
PowerHot swappable battery with approx. 3 hrs operation
Battery recharge time2 hrs
EnclosureIP65, NEMA4, GMP
Dimensions complete system230 x 210 x 150 (+50)1)mm / 9 x 8.3 x 5.9 (+2*) inch
1)­ length of process flange
Weight6.5 kg / 14.3 lbs
Measurement triggerInternal software controlled 3D position measurement
CommunicationsWireless LAN (802.11b/g)
SoftwareWindows XP Embedded
NovaPac® / NovaMath(21CFR part11 compliant)
ValidationSoftware features for validation routines according OQ/PQ
Accessories4” hygienic tri-clamp connection
Sample adapter with a sapphire window
Start up and training
NIST traceable reflectance standards for system validation
IQ/OQ documentation
Service contracts